Troubleshooting EMEC controllers & Alarms

Here are some quick troubleshooting tips to help you figure out what is going on with your EMEC controller.

Alarms

Alarm Types

Out of Range Alarms: The controller number is outside the minimum and maximum set points.

Dosing Alarms: One of the chemical pumps has reached their run time limit

How to clear a dosing alarm or any other alarm:

1) Login to controller
2) Exit back to main screen
3) Contact VivoAquatics support if this happens regularly.

Troubleshooting

Controller Offline

Power cycle needed

Chlorine not dosing/ Chlorine Low

Chlorine Barrel/Container is empty

Chlorine injector is clogged

Probe Failure

Setpoints not set correctly

Stenner needs a new tube

Dosing limit to low

Power cycle needed

Acid not dosing/ pH High

Acid Barrel is empty

Probe Failure

Setpoints not set Correctly 

Stenner tube needs replaced

Dosing limit to low

Power cycle needed

Probe Failure

Controller not reading the correct numbers

Probe Unresponsive

Probe needs cleaning

Probes needs replaced

Power Cycle needed

Need to change cap & gel (ppm only)

Chlorine too high

Setpoints not set correctly

Wrong feed mode set consider Proportional PWN 

Too much chlorine going through Stenner- turn down percentage

Dosing limit too high

Adding chemicals by hand

pH too low

Setpoints not set correctly

Wrong feed mode set consider Proportional PWN

Too much acid going through Stenner- turn down percentage

Dosing limit too high

Adding chemicals by hand.

Alkalinity not within limit

No Flow

Adjust sample flow valves to allow more flow

Backwash/clean filter if necessary

Clean controller pre-filter

All these can be done on site with the controller. If you are still having problems- a replacement part may be needed. 

*Remember automation equipment needs regular supervision and maintenance in order to be successful.